Thermocouple Open Circuit Test Procedure
Purpose
This procedure outlines the steps to test a thermocouple's voltage output using a multimeter to determine if it is functioning properly.
Required Tools and Equipment
-
Multimeter (set to millivolts DC)
-
3/8” open-end wrench
-
Alligator clip leads
Procedure
-
Preparation
-
Ensure the fireplace or appliance is off and cool before starting.
-
Identify the thermocouple connection point on the gas valve.
-
-
Disconnect the Thermocouple
-
Using a 3/8” open-end wrench, carefully unthread the thermocouple connector nut from the gas valve.
-
-
Connect the Multimeter
-
Attach one multimeter lead to the positive contact at the end of the thermocouple.
-
Attach the second multimeter lead to the copper casing of the thermocouple using alligator clips.
-
-
Set the Multimeter
-
Ensure your multimeter is set to the appropriate millivolts DC (mVDC) range.
-
-
Ignite the Pilot Flame
-
Light the pilot flame and hold in the control knob to maintain the flame.
-
Continue to hold the control knob until the multimeter reading stabilizes (approximately 45 seconds).
-
-
Observe the Reading
-
A healthy thermocouple should produce a reading of 18 to 30 mVDC.
-
The reading may display as positive or negative, depending on polarity — this is normal and does not indicate a fault.
-
-
Troubleshooting Low Readings
-
If the reading is below 18 mVDC, inspect the pilot flame:
-
Ensure the pilot flame is properly adjusted and enveloping the thermocouple tip.
-
Clean or adjust the pilot assembly if necessary before replacing the thermocouple.
-
-
Test Results
| Test Step | Pass/Fail | Notes |
|---|---|---|
| Multimeter connected properly | ||
| Stable voltage reading obtained | ||
| Voltage reading within 18-30 mVDC | ||
| Pilot flame inspection performed (if necessary) |
Conclusion
-
If the thermocouple voltage falls below the acceptable range and the pilot flame is confirmed to be correct, replace the thermocouple.
-
If the voltage is within range, no replacement is required.


